The equipment of clamp type multi product multimeter calibrator is
compliance with national verification regulation: JJG124-2005
“Ammeter, voltmeter, power meter and resistance meter verification
rules” and related national standards. 0.05 single-phase AC/DC
standard source, can verify magnitude 0.2 and the following AC/DC
voltmeter, ammeter, frequency meter, resistance meter. The source
signals by using DSP and 16-bit high-speed analog-to-digital
converters of controllable sine wave, the distorted wave signal
source.The equipment comes with RS-232 interface, can be connected
with PC to a semi-automatic verification system. It has precise
interface and multi-functions including verification, storage and
query.The equipment adopts color LCD screen, chart character
display is clear, with high precision, stable and reliable,
convenient operation and flexible characteristics.
Theory
Calibration of multimeter in measurement technology and metrology
is the comparison of measurement values delivered by a device under
test with those of a calibration standard of known accuracy. Such a
standard could be another measurement device of known accuracy, a
device generating the quantity to be measured such as a voltage, or
a physical artefact, such as a metre ruler.
The outcome of the comparison can result in no significant error
being noted on the device under test, a significant error being
noted but no adjustment made, or an adjustment made to correct the
error to an acceptable level. Strictly speaking, the term
calibration means just the act of comparison, and does not include
any subsequent adjustment.
The calibration standard of multimeter and clamp meter is normally
traceable to a national standard held by a National Metrological
Institute.
Electrical parameters | ||
---|---|---|
Power supply | Single phase AC 220V±10% or 110V±10%, 50/60HzAC | |
AC Voltage output | ||
range | 200mV, 400mV, 2000mV, 4000mV, 20V, 100V, 200V, 500V, 1000V | |
Range 20V-1000V | Accuracy | ±(0.03%RD+0.02%FS) |
Stability | ≤0.01% FS /60s | |
Range 200mV-4000mV | Accuracy | ±(0.3%RD+0.2%FS) |
Stability | ≤0.04% FS /60s | |
Load capacity | 20VA | |
Waveform distortion | ≤0.3% | |
Adjustable Range | 0-120% FS (1000V not included) | |
Fineness | 5×10-5 | |
AC Current Output | ||
Range | 2mA, 4mA, 20mA, 40mA, 0.2A, 0.5A, 2A, 5A, 10A, 20A | |
Range 0.2A-20A | Accuracy | ±(0.03%RD+0.02%FS) |
Stability | ≤0.01% FS /60s | |
Range 2mA-40mA | Accuracy | ±(0.3%RD+0.2%FS) |
Stability | ≤0.04% FS /60s | |
Load capacity | 20VA | |
Waveform distortion | ≤0.3% | |
Adjustable Range | 0-120% FS | |
Fineness | 5×10-5 | |
Frequency Output | ||
Range | 45-65Hz | |
Fineness | 0.001Hz | |
Accuracy | 0.01Hz | |
Harmonic Wave | ||
Times | 2-31 | |
Extent | 0-20% | |
Each harmonic phase fineness | 0.01·N (N is harmonic order) | |
DC voltage output | ||
Range | 200mV, 400mV, 2000mV, 4000mV, 20V, 40V, 100V, 200V, 400V, 1000V | |
Load capacity | 20W | |
Adjustable range | 0-110% FS (1000V not included) | |
Adjustable fineness | 5×10-5 | |
Stability | ≤0.01% FS /60s (peak-peak value) | |
Accuracy | ±(0.03%RD+0.02%FS) | |
Ripple wave | ≤1% FS | |
DC current output | ||
Range | 20uA, 40uA, 200uA, 400uA, 2mA, 4mA, 20mA, 40mA, 0.2A, 1A, 2.5A, 5A, 10A, 20A | |
Load capacity | 20W | |
Adjustable Range | 0-110% FS (1000V not included) | |
Adjustable Fineness | 5×10-5 | |
Stability | ≤0.01% FS /60s (peak-peak value) | |
Accuracy | ±(0.06%RD+0.04%FS) | |
Ripple wave | ≤1% FS | |
Resistance box | ||
Range | 0-200Ω, 200-400Ω, 0-2kΩ, 2kΩ-4kΩ, 0-20kΩ, 20kΩ-40kΩ, 0-200kΩ, 200kΩ-400kΩ, 0-2MΩ, 2MΩ-4MΩ | |
Fineness | 0.02% | |
Range 0-2kΩ, 0-20kΩ, 0-200kΩ | ||
Range 0-2kΩ | Adjustable Range | 0-100%FS |
Accuracy | ±(0.12%RD+0.08%FS) | |
Range 0-200Ω,0-2MΩ | Adjustable Range | 0-100% FS |
Accuracy | ±(0.3%RD+0.2%FS) | |
Range 2kΩ-4kΩ, 20kΩ-40kΩ, 200kΩ-400kΩ | ||
Range 2kΩ-4kΩ | Adjustable Range | 5%-100%FS |
Accuracy | ±(0.12%RD+0.08%FS) | |
Range 200-400Ω, 2MΩ-4MΩ | Adjustable Range | 5%-100% FS |
Accuracy | ±(0.3%RD+0.2%FS) | |
Clamp meter range | ||
200A | Accuracy | ±0.2% |
600A | Accuracy | ±0.5% |
1000A | Accuracy | ±1% |
Dot frequency square wave output | ||
Range | 1kHz, 5kHz, 10kHz, 50kHz, 100kHz, 200kHz, 1000kHz | |
Square wave amplitude | 4.5V | |
Setting accuracy | 5×10-5 | |
Mechanical parameters | ||
Dimension (L×W×H) (mm) | 440x360x160 | |
Weight (kg) | 12 | |
Environmental conditions | ||
Operating temperature | 20°C±10°C | |
Relative humidity | ≤85%RH |
Beijing GFUVE Instrument Transformer Manufacturer Co.,Ltd.
Beijing GFUVE Instrument Transformer Manufacturer Co., Ltd. was established in 2008. It combined design, manufacturing and sales together as one high-tech enterprise. GFUVE Instrument Transformer company mainly produces less than 35kV current transformer and voltage transformer, split core ct and current clamp models etc.
GFUVE is a leader manufacturer of split core current transformer products in China. It is professional in current transformer, potential transformer and related accessory. The product of outdoor split core current transformer develops very quickly in the global market. Our company has achieved certification to ISO9001. The products are sold to over 30 provinces in China and exported to overseas including South America, Middle East, Southeast Asia, Australia and Africa.